Calibrating the 920i – Rice Lake RailBoss™ Rail Scales - Calibration Guide User Manual

Page 3

Advertising
background image

Railboss Calibration Guide

3

Use the following steps to properly trim the JB4SP junction box.

1. Make sure jumpers are in place to enable

trimming of the cells corresponding to each
load cell. See Figure 4 for the location of
jumpers JP1, JP2, JP3, and JP4.

2. Set all potentiometers fully clockwise to give

maximum signal output from each cell (see
Figure 4 for location of potentiometers)

3. Zero the indicator and place the RailBoss

Calibrator over each load cell in turn. The
amount of test weight is recommended to be
30,000 lb.

4. Record the value displayed on the indicator

after the calibrator is placed in turn on each
load cell. Allow the scale to return to zero each
time to check for friction or other mechanical
problems. Select the load cell which has the
lowest value as your reference point. This cell
will not be trimmed.

5. Replace the same test load over each cell in

turn. Using the corresponding potentiometer,
trim each cell down to equal the reference load

cell. check all cells again for repeatability and,
if necessary, repeat Steps 3 and 4.

6. Pull excess cable out of the enclosure and

tighten the cord grip assemblies with a wrench.
To be watertight, each cord grip must be
tightened so the rubber sleeve begins to
protrude from the hub.

7. Unused hubs must be properly plugged to

prevent moisture entry. See Rice Lake’s
Electronic Replacement Parts Catalog to order
extra hole plugs.

8. Remove the desiccant from the plastic bag, and

insert the desiccant bag into the junction box
before closing. Inspect the desiccant during
normal service and change as needed.

9. Replace the cover and torque the cover screws

in an alternating pattern. Torque to 15 in/lb to
be certain the gasket is compressed equally in
all locations.

Calibrating the 920i

The

920i

can be calibrated using the front panel, serial commands, or

iRev

. Each method consists of:

Zero calibration

Entering the test weight value

Span calibration

Figure 5. Calibration (CALIBR) Submenu

SCALES

SERIAL

FEATURE

PFORMT

SETPTS

DIG I/O

VERS

SCALE 1

CALIBR

WZERO

WVAL

WSPAN

WLIN

REZERO

POINT 4

POINT 5

POINT 1

POINT 2

POINT 3