Irev calibration – Rice Lake RailBoss Rail Scales - Calibration Guide User Manual
Page 5
Railboss Calibration Guide
5
iRev Calibration
The
iRev
Calibration Wizard provides step-by-step
scale calibration. With the
920i
connected to a PC,
select the Calibration Wizard from the Tools menu on
the
iRev
Scales display, then follow the steps listed
below to calibrate the scale.
1. On the first Calibration Wizard display (see
Figure 7), select Standard Calibration. Click
the
Next
button to continue.
Figure 7.
iRev Calibration Wizard
2. Select the scale to calibrate (Scale #1).
Figure 8.
iRev Scale Selection Display
3. Enter the RailBoss Calibrator load value used
to calibrate the scale (30,000 lb).
Figure 9.
iRev Test Weight Value Display
4. Ensure the scale load is at zero. Press the
Calibrate Zero
button to perform the zero
calibration. A message appears when the
process is complete.
Figure 10.
iRev Zero Calibration Display
5. Using the RailBoss Calibrator, apply the 30000
load to the scale. Press the
Calibrate Span
button
to perform the span calibration. A message
appears when the process is complete.
Figure 11.
iRev Span Calibration Display
6. Review the new calibration values then click
Finish
to close the Calibration Wizard. To
restore the current calibration values, click
Cancel
.