Irev calibration – Rice Lake RailBoss™ Rail Scales - Calibration Guide User Manual

Page 5

Advertising
background image

Railboss Calibration Guide

5

iRev Calibration

The

iRev

Calibration Wizard provides step-by-step

scale calibration. With the

920i

connected to a PC,

select the Calibration Wizard from the Tools menu on
the

iRev

Scales display, then follow the steps listed

below to calibrate the scale.

1. On the first Calibration Wizard display (see

Figure 7), select Standard Calibration. Click
the

Next

button to continue.

Figure 7.

iRev Calibration Wizard

2. Select the scale to calibrate (Scale #1).

Figure 8.

iRev Scale Selection Display

3. Enter the RailBoss Calibrator load value used

to calibrate the scale (30,000 lb).

Figure 9.

iRev Test Weight Value Display

4. Ensure the scale load is at zero. Press the

Calibrate Zero

button to perform the zero

calibration. A message appears when the
process is complete.

Figure 10.

iRev Zero Calibration Display

5. Using the RailBoss Calibrator, apply the 30000

load to the scale. Press the

Calibrate Span

button

to perform the span calibration. A message
appears when the process is complete.

Figure 11.

iRev Span Calibration Display

6. Review the new calibration values then click

Finish

to close the Calibration Wizard. To

restore the current calibration values, click

Cancel

.