3 configuration of dip switches, Configuration of dip switches – Rice Lake Remote Displays/Scoreboards - 234HE User Manual
Page 5
4
Switch #1--Data Transfer Rate
(F 1.5 Series EPROM Version)
(Series 1, 2, or 3 EPROM Version)
off
1200 BAUD
off
300 BAUD
on
2400 BAUD
on
1200 BAUD
Switch #2--Number of Data Bits
off
7 data bits
on
8 data bits
Switch #3--Number of Stop Bits
off
1 stop bit
on
2 stop bits
Switch #4--Odd Parity
off
no parity or parity disabled
on
odd / even parity enabled
Switch #5 and #6--Serial Input Format
5 off
6 on 20 mA Current Loop
5 on
6 off RS232C
Switch #7--Time Display (Unavailable Option)
off
time display disabled
on
time display enabled
Switch #8--Temperature Display (Unavailalble Option)
off
temperature display disabled
on
temperature display enabled
3.
Configuration of Dip Switches
The purpose of dip switches #1-#6 is to ensure that the remote display and the indicator
are communicating in the same format. Consult your indicator manual to find the indicator's
current configuration. Check that the correct EPROM has been installed. Disconnect power
and set switches 1-6. Switches #7 and #8 enable the time and temperature display if those
options are installed.