Rice Lake RS-130 - Calibration Instructions User Manual

Rs-130 retail scale calibration instruction sheet

Advertising
background image

85773 - Rev A

RS-130 Retail Scale Calibration Instruction Sheet

Authorized distributors and their

employees can view or download this

manual from the Rice Lake Weighing

Systems retail distributor site at

www.ricelakeretail.com

.

Warning

Procedures described require work

inside the scale enclosure. These

procedures are to be performed by

qualified service personnel only.

Weight Calibration

Follow the steps below to complete the calibration process.

1. Turn on power and press and hold the

CHANGE

key until the WEIGHT display shows

6.3

. Release the

CHANGE

key.

2. Press and hold the

ZERO

key until the WEIGHT display shows

SEtUP

.

3. Press the

ZERO

key to enter calibration mode. WEIGHT display flashes

CAL-1

and shows

UNIT 1

.

NOTE:

The WEIGHT display will not flash

CAL-1

if jumper JP1 is in the disable position. See “Setting JP1 to the

Enable/Disable Position” on page 2. if the WEIGHT display will not flash

CAL-1

.

4. Press the

CLEAR

key to select the weight unit (default is kg) to be used for calibration.

Unit 1 - kg (annunciator above kg lit in the UNIT PRICE $ display)

Unit 2 - lb (annunciator above lb lit in the UNIT PRICE $ display)

Unit 3 - oz (annunciator above oz lit in the UNIT PRICE $ display) - not available in Canada

5. Press the

CHANGE

key to accept the weight unit. The WEIGHT display shows:

00.000 (kg)

000.00 (lb)

0000.0 (oz) - not available in Canada

6. Press the

key to shift and the

CLEAR

key to modify the flashing digit. For example, If you use a 30lb

standard weight to calibrate; use the

key to shift the flashing digit to

000.00

and use the

CLEAR

key to

set the digit to

030.00

.

7. Press the

CHANGE

key and wait for the WEIGHT display to show

000000

and the UNIT PRICE $

display to show the internal offset value at (

xxxxx

.)

8. Place selected weight from step six on the platter. WEIGHT display shows the internal span value and

UNIT PRICE $ display shows the internal offset value at (

xxxxx

.)

9. Wait for the span value to become stable, then press the

CHANGE

key. Wait for the WEIGHT display to

flash

good

(twice), quickly remove the weight from the platter. After automatic restart, finish calibration.

10. Disable JP1 to maintain NTEP and Measurement Canada approval See “Setting JP1 to the Enable/