Troubleshooting – Rice Lake RLC Stainless Steel User Manual

Page 8

Advertising
background image

6

1.

Check weigh module for debris restricting load cell movement or
debris between scale and structure.

2.

Check that tank/vessel and modules are plumb, level and square at the
critical areas.

3.

Check all piping and conduit for connections which restrict vessel
movement.

4.

If check rods are used, loosen all connections to finger tight only for
testing.

5.

Check load cell cables for physical or water damage.

6.

Check all electrical connections, especially in the junction box.

If the problem still is not found:

7.

Check possible indicator malfunction by using a load cell simulator to
input a known good signal into the indicator.

8.

Disconnect each load cell’s signal leads at the junction box and check
individual load cell outputs with a multimeter. Then check input/
output impedances for comparison with load cell manufacturer’s
specifications.

If after all these checks the problem still cannot be isolated, reconnect all but one
load cell. Replace the load cell with a load cell simulator. Alternate so that each cell
is individually disconnected and replaced with a simulator. If there is a problem
with a particular load cell, the symptom should disappear when that load cell is
disconnected and replaced with the simulator.

Possible Cause

No return to zero

Mechanical binding or debris in seals or under load cells; may have lost system
calibration

Non-linearity

Thermal expansion or deflection under load causing binding or side load

Non-repeatability

Loose load cell mount; drifting caused by moisture, load cell overload or shock
damage; mechanical binding

Lost calibration

Out of level or plumb; moisture problem; mechanical binding

Drifting readout

Moisture in junction box, cables, or load cell; mechanical binding

Symptom

5. Troubleshooting

If the system powers up and gives some type of stable digital readout that varies with
the load on the system, any system problems are probably caused by factors other
than the load cells. The load cells are often blamed for a malfunctioning system, but
90% of the time, the problem lies elsewhere. Look for mechanical causes for your
problem first.

If the system can be calibrated but doesn’t return to zero, loses calibration, or
demonstrates non-linearity or non-repeatability, see the following chart for possible
causes and do the following checks.

Advertising