Clear/exit totalize, Specific gravity (on certain models) – Rice Lake Scout Pro Series Portable, Ohaus - Instruction Manual User Manual

Page 25

Advertising
background image

SCOUT

PRO

EN-21

Clear/Exit Totalize

Performing this next step will erase all totalized memory.

3

Y

Specific Gravity (on certain models)

Specific gravity is

enabled only when SPEC.Gr is turned ON in the Mode submenu in

the .S.E.T.U.P. menu. The Specific Gravity mode allows calculating the specific
gravity of a sample. SPEC.GR mode must be turned ON before removing platform
and installing hook.
Prepare the balance as shown. Remove the balance platform,
insert the hook and install the beaker stand.

Beaker is not supplied.

The sample is weighed in air
suspended from the hook and then
weighed in water. The formula is

dry wt - wet wt.

Specific Gravity Setups

dry wt

Top View Side View

1

(See note)

Place weight on hook

1

(With print on, air weight is printed)

Place weight in water

1

(With print on, wet weight is printed)

Specific Gravity Value
NOTE: “G” indicates a specific gravity value.
Specific Gravity has no unit of measure.

Beaker Stand

Advertising