Error Codes List

When internal and some external problems occur with the balance, the internal software
will display messages as error codes. Explanations are given for possible problems.
Try remedies in order indicated.

Error Codes

Err 1 chEct

Invalid checksum data

a) Cycle balance ON/OFF,
b) Return balance for servicing.

Err 2 LoAd

Overload (>cap+90d) or Underload

a) Remove load,
b) Check shipping lock,
c) Recalibrate.

Err 4 WEIght Incorrect calibration weight

a) Use correct calibration weights,
b) Check environmental conditions.

Err 5 999999 Displayed value >99999

a) Exit application,
b) Rezero balance.

Err 6 count

Parts counting error - balance exits parts counting

a) Exit application,
b) Rezero balance.

Err 8 FULL

RS232 buffer is full

a) Cycle ON/OFF,
b) Check RS-232 settings in balance and computer.

Err 9 dAtA

Internal data error. The next button press or a 5 second time-out
causes the balance to turn off

a) Cycle ON/OFF,
b) Return balance for servicing.