5. Function setting for extended battery run time
The battery run time can be extended by changing the function settings.
Item Setting
Description
Interface
6.. 1..F..
0 Stops input/output
☆
1 Numeric 6-digit format
2 Numeric 7-digit format
3 Extended numeric 7-digit format
Auto power off
9. A.P.
☆
0 Disable
1 Enable
Auto backlight off
A.. A..b..
☆
0 Disable
1 Enable
Backlight control
o.. b..L..
0 Off
1 On
☆
2 On when the balance is AC-adapter
operated, and off when the balance is
battery-operated
Stars (✩) indicate factory defaults.
[Interface]
When no outputs are to be made, we recommend that the interface function be set to
“
6. 1.F.O.” Setting the interface function to “6. 1.F.O.” deactivates the circuitry for
outputs, reducing battery consumption.
[Auto power off]
The balance automatically turns off when the balance is stable for about 5 minutes.
The balance automatically turns off when you forget to turn it off, extending the
battery life.
[Auto backlight off]
The backlight automatically turns off when the balance is stable for about 3 minutes.
When any key is pressed or any change in weight is detected, the backlight turns on.
This function is enabled when “
o.. b..L..1” is selected.
[Backlight control]
“
o.. b..L..2
” is selected when the balance is shipped. The backlight automatically turns
off when the balance is battery-operated.
-2-