4 aligning the scale, Weighing, 1 adjusting the display – Seca 762 User Manual

Page 35

Advertising
background image

Weighing • 35

English

1. Position the transport lock as shown in the adjacent

drawing.

2. Attach the transport lock to the lower section of the

housing.

5.4

Aligning the scale

ATTENTION!

Incorrect measurement due to force shunts

Soft floor coverings such as carpets may cause
weights to be measured incorrectly.
– Set the scale up so that only its feet are in

contact with the floor.



Place the scale on a firm, flat surface.

6. WEIGHING

CAUTION!

Bodily injury

Perform a function check before every use as
described in the section “Function check” on
page 38.

6.1

Adjusting the display

To obtain meaningful measured values you must check
before every use that the pointer of the scale reads “0”.

1. Ensure that the scale has no load.

2. Check the current position of the pointer.

3. Move the adjusting wheel so that the pointer reads

“0”.

– Turn clockwise: Pointer moves counter-

clockwise

– Turn counter-clockwise: Pointer moves

clockwise

Advertising
This manual is related to the following products:

760, 750, 761