What do I do if …? • 39

English

If you find faults or deviations during the function
check, first try to remedy the fault with the aid of the
section “What do I do if …?”.

CAUTION!

Bodily injury

If you find faults or deviations during the
function check which you are unable to remedy
with the aid of the section “What do I do if …?”,
you may not use the device.
– Have the device repaired by seca service or

an authorized service partner.

– You should also follow the instructions in the

section “Maintenance/Recalibration” on
page 40.

9. WHAT DO I DO IF …?

Malfunction

Cause/Remedy

... the scale does not
show any weight?

The device is not ready for use

- Remove transport lock

... the scale does not
show any plausible
weight?

• Before weighing pointer does not read “0”

- Adjust display

• Force shunt: scale is e.g. touching a wall or

patient is supporting his/her weight against the
wall.
- Move scale away from wall
- Use transport lock as wall spacer
- Ask patient to stand on the scale unaided