Defibrillator Synchronization

These instructions are not intended to replace existing hospital procedures relative to the provision of
cardiac electrical therapy and operation of the specific models of defibrillators. Use all safety standards
and clinical protocols as defined by your institution.

The Defibrillator Synchronization feature is designed to operate only with the Physio-
Control LIFEPAK 5

®

and LIFEPAK 6s

®

defibrillators.

Defib Sync Connector and Cables

The DEFIB SYNCHRO connector is on the Propaq’s right side panel.

Installing the Defib Sync Interface to the LIFEPAK 5

The Propaq Encore Defib Sync interface provides signal transmission between the Propaq
Encore and the LIFEPAK 5. The interface connects to the LIFEPAK 5 along the left side of the
defibrillator. Use the following instructions to install and remove the interface.

1.

Before installing the interface, check that
the contacts on the left side of the
LIFEPAK 5 are clean in order to ensure
signal transmission between the
LIFEPAK 5 and the Propaq.

Warning Use only the correct Welch Allyn cable with the LIFEPAK 5 or
LIFEPAK 6s. (These cables contain circuitry in addition to their wiring.) The
use of any other cable will result in incorrect operation. The Welch Allyn
Products and Accessories Booklet lists correct cables for use with Welch Allyn
products.

Defibrillator Application

Welch Allyn Part Number

LIFEPAK 6s

008-0154-00

LIFEPAK 5

008-0136-XX

Contact

Points