In-service simulated values, In-service values – Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 160

Advertising
background image

In-Service Simulated Values

Welch Allyn

160

Propaq Encore Reference Guide

In-Service Simulated Values

In-Service Values

Channel

Display

Initial Value

Alternate Value

ECG

Waveform

Normal sinus rhythm, 1mV,
Lead II

Normal sinus rhythm

ECG

Heart Rate

80 beats per minute

125 beats per minute

RESP

Respiration Rate

12 breaths/minute

31 breaths/minute

P1

Waveform

Arterial

Same as Initial Value

P1

Pulse Rate

80 pulses per min

125 pulses per min

P1

Systolic

121 mmHg

120 mmHg

P1

Diastolic

79 mmHg

85 mmHg

P1

Mean

96 mmHg

103 mmHg

P2

Waveform

Pulmonary Artery

Same as Initial Value

P2

Pulse Rate

80 pulses per min

125 pulses per min

P2

Systolic

25 mmHg

25 mmHg

P2

Diastolic

9 mmHg

12 mmHg

P2

Mean

15 mmHg

18 mmHg

NIBP

Mode

Manual (Auto cancels in-
service)

Same as Initial Value

NIBP

Numerics

Actual values from patient

Actual values from patient

T1

Numeric

37.0° C

39.1° C

T2

Numeric

36.4° C

37.4° C

∆T

Numeric

0.6° C

1.7° C

SpO

2

Waveform

Normal, 2x

Same as Initial Value

SpO

2

Rate

80 pulses per min

125 pulses per min

SpO

2

Numeric

97%

88%

CO

2

Waveform

Normal

Hyperventilating

CO

2

ETCO

2

Numeric

38 mmHg

60 mmHg

CO

2

INCO

2

Numeric

0 mmHg

8 mmHg

CO

2

Breath Rate

12 breaths/minute

31 breaths/minute