Chapter 4 Reviewing the ECG Settings

Welch Allyn CP 100 Electrocardiograph

Item (in

Figure 21

on page 33)

Description

A. Patient data

See

“Reviewing the Patient Data Fields Available”

on page 37.

B. ECG measurements

Standard.

C. Interpretation (optional)

See

“Reviewing the Interpretation and Copy Settings for Auto Reports”

on

page 36.

D. Report status label

See

“Reviewing the Interpretation and Copy Settings for Auto Reports”

on

page 36.

E. 3 rows, 4 columns

See

“Reviewing the Format Settings for Auto Reports”

on page 35.

F.

Rhythm leads

See

“Reviewing the Format Settings for Auto Reports”

on page 35.

G. Software version

See also

“Reviewing the Device Information”

on page 29.

H. Device ID

See

“Device ID”

on page 27.

I.

Practice ID

See

“Practice ID”

on page 27.

J. Date and time

See

“Set Date/Time”

on page 27.

K. AC filter

See

“Mains Filter”

on page 40.

L.

Frequency range

Lower limit: baseline filter on = 0.5, off = 0.3
Upper limit: muscle filter on = 35, off = 150
See

“Adjust Baseline Filter”

on page 55 and

“Adjust Muscle Filter”

on page 55.

M Gain

See

“Adjust Gain”

on page 55.

N. Paper speed

See

“Adjust Speed”

on page 55.

O. Calibration pulse

Amplitude reference — represents the current height of a one-millivolt signal.
Is adjusted for the selected gain:

5 mm/mV = 0.5 x
10 mm/mV = 1 x
20 mm/mV = 2 x