Directions for Use

Chapter 5 Performing ECG Tests

47

4.

If the electrocardiograph’s display is blank, press

.

5.

If prompted to select a lead set type, select Standard or Augmented.
For a description of the augmented pediatric lead set, see

page 4

.

6.

If the Lead Status screen appears, as shown here, reattach any leads that are
flashing.

Figure 29. “Lead Status” Screen

The term “Artifact” might also appear on this screen. To eliminate artifact, you might
need to re-prep the patient, use fresh electrodes, or minimize patient motion during
the recording.

The most common ECG problems are poor electrode contact and loose leads.

When all leads have been connected for several seconds, the following screen
appears.

Figure 30. “Ready” Screen

7.

Go to the procedure for the type of ECG test you want to perform.

“Recording an Auto ECG”

on page 48

“Recording a Rhythm ECG”

on page 54

9:17AM Oct 16 08

Lead Status

Check lead

C1
C2

Press Auto ECG or Rhythm ECG to

9:17AM Oct 16 08

start

HR: 75