Specifications

Duet CO

2

Module Specifications

PHYSICAL SPECIFICATIONS

Size

3.15 in

× 3.9 in (80 mm × 100 mm)

Clearance

1.0 in (25.4 mm) above board
0.2 in (5.0 mm) below board

Orientation

No limitations

Weight

< 1 lb (454 gm) with both sticks and full pneumatics

SYSTEM PERFORMANCE SPECIFICATIONS

CO

2

Concentration Display Rng.

0 – 99 mmHg

Respiratory Rate Range

0 – 150 bpm

Typical Accuracy

a

+/- 2 mmHg,

0 – 40 mmHg

+/- 5% of reading

41 – 76 mmHg

+/- 10% of reading

77 – 99 mmHg

Start up Time

< 80 sec in mainstream mode from 25

o

C ambient; (5 W to

sensor heater)
< 30 sec typical in sidestream mode

Calibration

No routine calibration required

Calibration Stability

< 1% outside accuracy spec. after 12 months of continuous use

Mainstream Response Time

100 msec (10% to 90%)

Sidestream Rise Time

240 msec (10% to 90%)

Sidestream Delay Time

1.12 sec max with 7 ft L; 0.055 in I.D.; sample line @ 175 ml/min

Sidestream Flow Rate Range

variable from 90 to 200 ml/min

Flow Rate Accuracy

-20, +15% of set value