Welch Allyn Duet CO2 Module - User Manual User Manual

Page 64

Advertising
background image

Duet CO

2

Module OEM Implementation Manual

Page 64

Confidential

Welch Allyn OEM Technologies

Adult/Pediatric Multi-Use Airway Adapter Specifications

PHYSICAL SPECIFICATIONS

Overall Dimensions

2.04"L.x 0.88"W x 1.14"H
(5.2cm L x 2.24cm W x 2.9cm H)

Contributory Dead Space

6cc (6ml)

Fittings:

Male O.D. 15mm, per ISO 5356-1 and ASTM F1054
Female I.D. 15mm, per ISO 5356-1 and ASTM F1054

Resistance:

See Figure C

Respiration Rate

0 to 150 BPM

Nominal Weight:

4.8g

Material:

Polysulfone resin, aluminum and sapphire

Pressure Leak Tested at:

15psi

Sterilizing Method:

Steam autoclave

Environmental Conditions

Ambient Shipping/Storage:

Temperature
Relative Humidity
Barometric Pressure

-4° to 140° F (-20° to 60° C)
15-95% RH, non-condensing
795mmHg to 375mmHg (-1300ft to 45000ft)

Ambient Operating Conditions:

Temperature
Relative Humidity
Barometric Pressure

50° to 140° F (10° to 60° C)
15%-95%RH, non-condensing
795mmHg to 375mmHg (-1300ft to 45000ft)

Note:

This airway adapter is designed for a maximum flow of 120 l/min. This airway adapter is
designed to be sterilized via steam autoclave only. Do not exceed 250° F (120° C)
(14.7PSIG). Designed to withstand a minimum of 20 sterilization cycles.

Pressure Drop

(cm H

2

0)

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

0

10

20

30

40

50

60

70

80

90

100

110 120

Flow (Liters/min)

Figure C

Average Pressure Drop Across the Adult Airway Adapter