Lamp Replacement

1.

Turn unit off and unplug the power cord from both the wall outlet and
the Stryker Light Source.

2.

Remove the lamp door from the bottom of the Stryker Light Source
by unscrewing the lamp access knob on the back of the lightbox.

3.

Disconnect the lamp connector assembly and remove the lamp from
the holding spring assembly.

4.

Replace with Stryker replacement lamp #09500 only. Do not touch
the lamp itself, or the interior reflective surface of the lamp. Skin oils
on the lamp will cause premature lamp failure. Hold lamp by the
outside of the reflector or the connector only. Remove any grease or
fingerprints with a clean cotton swab moistened with alcohol. Do not
leave any lint on the lamp.

CAUTION: Lamp is hot. Allow the lamp to cool before
replacing.