1-13

Maintenance And Repair

For minor trouble, refer to the troubleshooting section in this manual for
possible causes and corrective action.

Only qualified personnel should make electrical inspections of the Stryker
Light Source.

Repair: To locate qualified personnel, contact your local authorized
Stryker distributor or call Stryker directly at 1-408-754-2000.

Repair must be performed by authorized personnel. Failure to do so
invalidates warranty.

fan is
operating.

The lamp was not
able to “cool down”
before restarting.

Wait at least 15 seconds after
turning the Light Source off
before turning it back on.

The Light Source
has overheated.

Verify that all ventilation slots
and the fan area are not
obstructed. Mount the Light
Source on the supplied
mounting bracket and verify
there is always at least 3-
inches clearance.

The lamp has burned
out.

Replace with a new Stryker
lamp. Refer to the lamp
replacement section.