SureTemp Plus Module OEM Implementation
OEM Implementation
Welch Allyn OEM Technologies
Confidential 19
Time-out
All time-outs return the module to the standby state. After three minutes in standby, the module
drops into the sleep state.
This table lists the time-outs for various conditions.
Table 3. Time-out
Time-out
Period Before
Transition
Next State
Power Reset
4 seconds
standby
Probe Connect
5 seconds
standby
Measurement Out-of-Range
Below 80
°
F, above 110
°
F
10 minutes
standby
Predict Completion
30 seconds
standby
Error
5 seconds
standby
Standby
3 minutes
sleep