Welch Allyn Connex Vital Signs Monitor Calibration/Verification Module Notes - Quick Reference Guide User Manual

Page 3

Advertising
background image

Species

Code Definition /

Description

Original / Alert

type

Traditional code

map

Details

Host

controller

action

5

18

Host controller

alert

Internal error

None

Request

appropriate

algorithm.

5

19

Host controller

alert

Internal error

None

Request

appropriate

code.

5

20

Host controller

alert

Internal error

None

Request

appropriate

patient code.

5

21

Host controller

alert

Internal error

None

Request

appropriate

configuration.

5

22

Equipment alarm E11

Cuff pressure

duration

exceeded.

Power down.

5

23

Equipment alarm C12

Device outside

operating

temperature

range.

Clear.

5

24

Equipment alarm C13

Low battery

level. Charge

battery.

Clear.

5

25

Equipment alarm C13

High battery

level.

Clear.

6

1

Equipment alarm Internal error

Internal NIBP

error. Contact

Welch Allyn

Technical

Service.

Power down.

6

2

Equipment alarm Internal error

Internal NIBP

error. Contact

Welch Allyn

Technical

Service.

Power down.

6

3

Equipment alarm Internal error

Internal NIBP

error. Contact

Welch Allyn

Technical

Service.

Power down.

6

4

Equipment alarm E10

Cuff pressure

limits exceeded.

Reset Mod PG.

6

5

Host controller

alert

C07

Internal NIBP

error.

Host controller

to retry Auto

cycle request.

Advertising