Welch Allyn Connex Vital Signs Monitor Calibration/Verification Module Notes - Quick Reference Guide User Manual

Page 6

Advertising
background image

Code Definition /

Description

Origin / Alert

type

Traditional code

map

Display message to

operator

Host controller

action

8

Too few steps.

Patient alert.

C05

Unable to determine

blood pressure. Check

connections; restrict

patient movement.

Make the error

clearable, and

allow another

cycle to be run.

9

A neonate cuff was

detected in adult

mode.

Operator alert. Not applicable.

Message displayed

depends on the host

controller environment.
Message could be due to

neonatal patient

movement.

Make the error

clearable, and

allow another

cycle to be run.
Change patient

mode or an

equipment alert

(C03).

10

Re-inflation too late. Patient alert.

C05

Unable to determine

blood pressure. Check

connections; restrict

patient movement.

Make the error

clearable, and

allow another

cycle to be run.

11

Too many re-

inflates.

Patient alert.

C05

Unable to determine

blood pressure. Check

connections; restrict

patient movement.

Make the error

clearable, and

allow another

cycle to be run.

12

Could not bleed

pressure below

SVRP.

Equipment

alert.

C02

Unable to release cuff

pressure.
Check tubing for kinks

and connection integrity.

Make the error

clearable, and

allow another

cycle to be run.

13

Air leak detected.

Equipment

alert.

C03

Inflation too quick. Check

for kinks in the tubing

and connections.

Make the error

clearable, and

allow another

cycle to be run.

14

Safety check failed. Equipment

alert.

C10

Safety fault detected

during cycle. Cycle

cancelled.

Power down.

Advertising