Invasive pressure messages

The following messages can appear in the numerics window.

NOT ZEROED (or NO ZERO). No zero reference has been established. The monitor
displays the pressure waveform, but to protect against erroneous readings, the
pressure waveform scale is not displayed. To remove this message, zero the
transducer.

ZEROING. This message briefly appears as the transducer is being zeroed.

ZEROED. This message appears after the zero value has been accepted. It remains
for eight seconds and is replaced by the current pressure values.

ZERO REJECTED (or REJECT). Unable to establish a zero reference value. The
message remains until the CANCEL button is pressed.

CANCELED. This message appears if CANCEL is pressed while the channel is
zeroing.

These equipment messages can appear in an equipment alert window.

TRANSDUCER NOT DETECTED. The transducer connection is broken.

TRANSDUCER SHORT CIRCUIT. This message appears when the Propaq Encore
senses a short in the transducer. The transducer should be replaced.

INCOMPATIBLE TRANSDUCER. Visit

www.welchallyn.com

to confirm you are using

a compatible transducer.