Printing nibp measurements, And resp waveforms – Welch Allyn Vital Signs Monitor, Propaq Encore - Quick Reference Guide User Manual

Page 86

Advertising
background image

80

Printing

Welch Allyn Propaq Encore Vital Signs Monitor

CO

2

and RESP waveforms

CO

2

waveforms are printed on a grid with major divisions (vertical dotted lines) every

5 mm, and the pressure scale grids are printed horizontally. The waveforms and numerics
are labeled as MCO

2

for mainstream CO

2

and SCO

2

for sidestream CO

2

. If neither CO

2

option is active, the label is simply CO2.

On the

SNAPSHOT

command only, the CO

2

and RESP waveforms will be printed with

different sweep speeds than the other waveforms.

Printing NIBP measurements

You can print the results of an NIBP measurement each time one occurs. This printout is
called the NIBP Ticket. The NIBP Ticket must be turned on in the Printer Setup window.

1.

To turn on NIBP TICKET, press SETUP, MORE, PRINTER. The printer setup window
appears.

2.

Use the NEXT and CHANGE buttons to select and turn on the NIBP TICKET.

This symbol indicates that the NIBP reading was taken in the presence of high motion artifact while monitoring ECG.
Artifact can affect accuracy. To help reduce artifact, see

“Place and connect the cuff”

on page 44.

Advertising