Welch Allyn ABCs of Vital Signs Monitor 300 Series - Quick Reference Guide User Manual

The abc’s of vital signs monitor 300 series

Advertising
background image

VSM 300 FAQ

MC 7722 09/2011

The ABC’s of Vital Signs Monitor 300 Series

A

Age Ranges

Neonatal

Pediatric

Adult


Neonate: Term birth (37 weeks) through 28 days, or up to 44 weeks gestational age.

Pediatric: 29 days through 12 years of age.

Adult: 13 years and older.

B


Battery Info

Uses a sealed lead-acid battery.
Provides 8 hours of continuous monitoring of Adult NIBP (3 minute intervals/165 readings)
with simultaneous temperature and continuous SpO2.
Charges to 90% of capacity in 12 hours.
Should be charging when not in use.
Patient data is erased if battery goes dead.
If not plugged in, monitor turns off 15 minutes after the battery discharge reaches “failure
level.”

Battery indicator is located on front panel when the unit is turned on. When battery is LOW,
a warning tone sounds and the low battery indicator flashes. (To avoid shutdown, plug in and
connect the AC power). When the low battery indicator appears, there is enough charge
remaining to perform at least ONE of the following functions over the next hour:

20 NIPB cycles

30 minutes of SpO2

20 temperatures

a single printing of 99 stored data sets

C

Cleaning

Device
Wipe the monitor with a cloth slightly dampened with warm water, a mild detergent or
appropriately diluted, non-staining disinfectant. Never wet or immerse the device.

Accessories
Do not immerse, soak, or autoclave. Wipe the NIBP hose and reusable cuffs with a damp
cloth moistened in a mild detergent. Clean the SpO2 sensor with a cloth dampened with
70% alcohol or 10% bleach solution. Wipe the temperature probe with a cloth dampened
with alcohol, warm water, or an appropriately diluted, non-staining disinfectant solution.
Clean removable probe well by swabbing with a cloth dampened with 70% isopropyl alcohol
or 10% bleach solution. Probe well can be immersed.

D


Defaults

To access the Internal Configuration menu, press and hold the MENU

button for 3

seconds. Continue to press the MENU button until you reach the setting of interest. Use the

UP and DOWN

buttons to change the values. Set the new value either by doing

nothing for 10 seconds or by pressing any button other than the UP and DOWN buttons.

If you press a function button such as the NIBP

button, the monitor returns to

normal operation with that function activated. Powering OFF

the device will save all

the values.
Note: When Time/Date are changed, patient data is erased.

E

End Monitoring

To shut off the device, press

. This will erase all stored patient data but not settings

or configurations.