I. introduction, Introduction – Brookfield DV-II+ Programmable Viscometer User Manual

Page 3

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 3

Manual No. M/97-164-F1 102

I.

INTRODUCTION

The Brookfield Programmable DV-II

+

Viscometer measures fluid viscosity at given shear rates.

Viscosity is a measure of a fluid’s resistance to flow. You will find a detailed description of the
mathematics of viscosity in the Brookfield publication ”More Solutions to Sticky Problems” a copy
of which was included with your DV-II

+

.

The principal of operation of the DV-II

+

is to drive a spindle (which is immersed in the test

fluid) through a calibrated spring. The viscous drag of the fluid against the spindle is measured
by the spring deflection. Spring deflection is measured with a rotary transducer. The measure-
ment range of a DV-II

+

(in centipoise or milliPascal seconds) is determined by the rotational

speed of the spindle, the size and shape of the spindle, the container the spindle is rotating in, and
the full scale torque of the calibrated spring.

There are four basic spring torque series offered by Brookfield:

Spring Torque

Model

dyne-cm

milli Newton - m

LVDV-II

+

673.7

0.0673

RVDV-II

+

7,187.0

0.7187

HADV-II

+

14,374.0

1.4374

HBDV-II

+

57,496.0

5.7496

The higher the torque calibration, the higher the measurement range. The measurement range for
each torque calibration may be found in Appendix B.

All units of measurement are displayed according to either the CGS system or the SI system.

1. Viscosity appears in units of centipoise (shown as “cP”) or milliPascal-seconds (shown

as “mPa•s”) on the DV-II

+

Viscometer display.

2. Shear Stress appears in units of dynes/square centimeter (“D/cm

2

”) or Newtons/square

meter (“N/m

2

”).

3. Shear Rate appears in units of reciprocal seconds (“1/SEC”).
4. Torque appears in units of dyne-centimeters or Newton-meters (shown as percent “%”

in both cases) on the DV-II

+

Viscometer display.

Note: To change CGS to SI units on the display - see Section III.2.2.

The equivalent units of measurement in the SI system are calculated using the following conversions:

SI

CGS

Viscosity:

1 mPa•s

=

1 cP

Shear Stress: 1 Newton/m

2

=

10 dyne/cm

2

Torque:

1 Newton-m

=

10

7

dyne-cm

References to viscosity throughout this manual are done in CGS units. The DV-II+ Viscometer
provides equivalent information in SI units.