Brookfield DV-III Rheometer User Manual

Page 57

Advertising
background image

Brookfield Engineering Laboratories, Inc.

Page 57

Manual No. M/98-211-A0701

Universal Spindle Ranges

Example 1:

Determine the full scale viscosity range of the LV3 spindle running on an RV series
Rheometer at 45 RPM.
Spindle Speed = 45 RPM
LV Spindle Range Coefficient for RV series Rheometer = 1,280,000
Full Scale Viscosity Range = 1,280,000 = 28,444 cP (mPa•s)

45

Example 2:

Determine the full scale viscosity range of the LV3 spindle running on a 2xHA
Rheometer at 103.5 RPM.
Spindle Speed = 103.5 RPM
LV3 Spindle Range Cooefficient for 2xHA Rheometer = 2 x 2,560,000
Full Scale Viscosity Rnage = 5,120,000 = 49,468 cP (mPa•s)

103.5

Note:

The maximum viscosity taht should be taken from the DV-III+ Rheometer is at 100%
of any full scale spindle/speed range.

DIN-81

81

3,420

36,500

73,000

292,000

DIN-82

82

3,420

36,500

73,000

292,000

DIN-83

83

11,340

121,300

242,600

970,400

DIN-85

85

1,144

12,200

24,400

97,600

DIN-86

86

3,420

36,500

73,000

292,000

SC4-14

14

117,200

1,250,000

2,500,000

10,000,000

SC4-15

15

46,880

500,000

1,000,000

4,000,000

SC4-16

16

120,000

1,280,000

2,560,000

10,240,000

SC4-18

18

3,000

32,000

64,000

256,000

SC4-21

21

4,688

50,000

100,000

400,000

SC4-25

25

480,000

5,120,000

10,240,000

40,960,000

SC4-27

27

23,4000

250,000

500,000

2,000,000

SC4-28

28

46,8880

500,000

1,000,000

4,000,000

SC4-29

29

93,750

1,000,000

2,000,000

8,000,000

SC4-31

31

30,000

320,000

640,000

2,560,000

SC4-34

34

60,000

640,000

1,280,000

5,120,000

SC4-37

37

23,4000

250,000

500,000

2,000,000

CP-40

40

307

3,270

6,540

26,160

CP-41

41

1,151

12,280

24,560

98,240

CP-42

42

600

6,400

12,800

51,200

CP-51

51

4,800

51,200

102,400

409,600

CP-52

52

9,216

98,300

196,600

786,400

Spindle

Entry Code

LV

RV

HA

HB

SPINDLE RANGE COEFFICIENT

Rheometer Series

Notes:

RV spindle ranges are calculated with the RV Guardleg in use.
LV spindle ranges are calculated with the LV Guardleg in use.

Advertising