Brookfield DV-III Rheometer User Manual

Page 68

Advertising
background image

Brookfield Engineering Laboratories, Inc.

Page 68

Manual No. M/98-211-A0701

Brookfield Viscosity Standard Fluid - General Information

We recommend that Brookfield Viscosity Standard Fluids be replaced on an annual basis, one year from
date of initial use. These fluids are pure silicone and are not subject to change over time. However,
exposure to outside contaminants through normal use requires replacement on an annual basis.
Contamination may occur by the introduction of solvent, standard of different viscosity or other foreign
material.

Viscosity Standard Fluids may be stored under normal laboratory conditions. Disposal should be in
accordance with state, local and federal regulations as specified on the material safety data sheet.

Brookfield Engineering Laboratories does not recertify Viscosity Standard Fluids. We will issue
duplicate copies of the Certificate of Calibration for any fluid within two years of the purchase date.

Brookfield Viscosity Standard Fluids are reusable provided they are not contaminated. Normal practice
for usage in a 600 ml beaker is to return the material from the beaker back into the bottle. When using
smaller volumes in accessories such as Small Sample Adapter, UL Adapter or Thermosel, the fluid is
normally discarded.

Calibration Procedure for LV(#1-4) and RV,HA,HB(#1-7) Brookfield spindles:

1) Place the viscosity standard fluid (in the proper container) into the water bath.

2) Lower the DV-III+ into measurement position (with guard leg if LV or RV series Rheometer

is used).

3) Attach the spindle to the Rheometer. If you are using a disk shaped spindle, avoid trapping air

bubbles beneath the disk by first immersing the spindle at an angle, and then connecting it to the
Rheometer.

4) The viscosity standard fluid, together with the spindle, should be immersed in the bath for a

minimum of 1 hour, stirring the fluid periodically, prior to taking measurements.

5) After 1 hour, check the temperature of the viscosity standard fluid with an accurate thermometer.

6) If the fluid is at test temperature (

+

/- 0.1

°

C of the specified temperature, normally 25

°

C),

measure the viscosity and record the Rheometer reading.

Note:

The spindle must rotate at least five (5) times before readings are taken.

7) The viscosity reading should equal the cP value on the viscosity fluid standard to within the

combined accuracies of the Rheometer and the standard (as discussed in the section entitled,
Interpretation of Calibration Test Results).

Calibration Procedure for a Small Sample Adapter

When a Small Sample Adapter is used, the water jacket is connected to the water bath and the water

is stabilized at the proper temperature:

1) Put the proper amount of viscosity standard fluid into the sample chamber. The amount varies

with each spindle/chamber combination. (Refer to the Small Sample Adapter instruction
manual.)

2) Place the sample chamber into the water jacket.

Advertising