Brookfield RST Rheometer User Manual

Page 16

Advertising
background image

Brookfield Engineering Labs., Inc.

page 16

Manual No. M14-223

II.4.5.1 Assembly of FTK Temperature Control Device (Water Jacket)

This section describes assembly and connection of the following measuring systems:

Figure II-2

Components:
1 Threaded joint for mounting flange

4 Hose conncetion for thermal oil discharge

2 Thermostatting chamber

5 Pt100 temperature sensor

3 Threaded joint for sample cup

6 Hose connction for thermal oil inflow


Optional temperature control devices available:

• Liquid temperature control device FTK for use in cylinder measuring systems in the

temperature range of -10°C ... +90°C (temperature control by liquid thermostat).

• If the optional KE cooling device is used, the temperature control device FTK can be

operated in the temperature range of -20°C ... +180°C.

Caution!

Temperature control device

The temperature control device FTK

without KE cooling unit must not be used

beyond the temperature range of -10°C ...

+90°C!

Figure II-3

Advertising