5 Putting into operation

64

R-220 SE Operation Manual, Version F

NOTICE

Risk of instrument damage by wrong mains supply.
• External mains supply must meet the voltage given on the type plate
• Check for sufficient grounding

5 .18

Adjusting the snap flange

System tightness is essential when working with a Rotavapor. Thus, the snap flange assembly should
be checked at every evaporating flask installation. This is necessary since every glassware has its
unique fitting accuracy.

Clockwise = increasing the tension
Counterclockwise = decreasing the tension

• When passing the top dead center of the closing mechanism a clear resistance must be felt.
• Moving the lever must never require excessive force!

➡ If the required lever force is too weak the flange may not be thight and will wear out the PTFE seal.

In this case, increase the tension of the closure.

➡ If the required lever force is too high (e.g. the lever can not be shifted over the top dead center)

reduce the tension of the closure.