8 Troubleshooting

81

R-220 SE Operation Manual, Version F

Error

Additional error indicator on display Possible cause

Remedy

E03 Drive unit icon flashes.

The rotary drive does not reach the
set sppec value.

• Drive defective or blocked.

• Press the “Aerate” button to confirm

the error message.

• Switch off the unit and check if

drive is blocked, rotate evaporating
flask manually.

Call service if problem is persistent.

E04 n.a.

Battery voltage too low.

• PB battery weak or defective.

• Press the “Aerate” button to confirm

the error message.

• Switch on the system over night to

allow recharging of the battery.

Call service if problem is persistent.

E05 n.a.

Hardware malfunction.

• Restart the unit.

Call service if problem is persistent.

E06 n.a.

Shortcut at aeration valve.

• Unplug valve and restart system for

causal research.

• Exchange valve if defective or run

evaporation with manual aeration.

Call service if problem is persistent.

E07 n.a.

Shortcut at cooling water valve.

• Unplug valve and restart system for

causal research.

• Exchange valve if defective or run

evaporation without cooling water
valve.

Call service if problem is persistent.

E08 n.a.

Shortcut at vacuum valve.

• Unplug valve and restart system for

causal research.

• Exchange valve if defective.

Call service if problem is persistent.

E11 ‘SHIELD OPEN’ message

appears.

Shield not closed at the time of
flask rotation start.

• Close bath shield.

• Press the “Aerate” button to confirm

the error message.

Call service if problem is persistent.