5 Putting into operation

17

C-660 Operation Manual, Version F

3

6

7

8

The collector head e is found on the collec-
tor arm h of the fraction collector. Clamp the
tubing guide f (metal pipe) tightly into the
collector head g.

Ensure that the height of the tubing guide f
is correct. It should be at least 1 cm above the
tube rim, so that it will not damage the tubing
during operation.

Now direct the FEP-tubing c from valve out-
let b and afterwards through the tubing guide
f, so that it pokes out approximately 3 mm.

Connect the valve to the WASTE VALVE con-
nection i.

9

Fig. 5.3: Installating the waste diverter valve

2 3 4

1

8

5

6

7

9 10

Fig. 5.4: Control arrangement on the Pump Controller

a

Power cord

b FUSE 0.5 AT: Fuse
c WASTE VALVE: Valve control
d WASTE LEVEL: Monitoring of valve bottle

level

e

TTL IN/OUT: TTL connection

f DETECTOR IN 1,2: Detector inputs 1 + 2
g X-T RECORDER 1,2: Chart recorder outlets

1 + 2

h RS 232: Series data interfaces
i

RS 485: 2× bus interface for peripheral
devices (fraction collector, pumps, etc.)

j

Address bus switch for remote control
position 1 or 2 for SepacoreControl,
position 0 for manual mode