5 Putting into operation

23

K-425 / K-436 Operation Manual, Version A

5 .3

Electrical connections

After the installation procedure has been completed successfully, the power plug of the SpeedDigester
must be connected to the mains for the digestion process.

Notice

Risk of instrument damage by wrong mains supply.

External mains supply must meet the voltage given on the type plate

Check for proper grounding

The used mains circuit has to:

provide the voltage that is given on the type plate of the instrument.

be able to handle the load of the connected instruments.

be equipped with adequat fusage and electrical safety measures, in particular proper earthing.

See also technical data of all components regarding the different minimum system requirements!

Note
Additional electrical safety measures such as residual current breakers may be necessary to meet
local laws and regulations! External connections and extension lines must be provided with an
earthed conductor lead (3-pole couplings, cord or plug equipment). All used power cords shall be
equipped with moulded plugs only to avoid risks due to unobservant defective wiring.