6 Operation

24

K-425 / K-436 Operation Manual, Version A

6

Operation

This chapter gives examples of typical instrument applications and instructions on how to operate the
instrument properly and safely. See also section 2.5 “Product safety” for general warnings.

6 .1

Operating controls and housing

The SpeedDigester K-425 / K-436 are equipped with thermostats. The table below lists the correlated
temperature (typical average values) to every heating level position of each thermostat.

Level

Set-point temperature in °C

Level

Set-point temperature °C

OFF

1

70

6

350

2

130

7

410

3

180

8

470

4

240

9*

520

5

300

10*

580

* Level 9 and 10 are highlighted red. On these high temperature levels the recondensation zone can

be displaced inside the safety zone, close to the suction module. This can cause ineffective digestion
and corrosion of sealings and glassware!

6 .1 .1

SpeedDigester K-425 — front side

Cooling chamber

Heating chamber

Heating indicator

Thermostat knob

Mains switch