8 .3

Reference measurement fails

When the deviation between the stored reference spectrum is out of limit, the reference measurement
will fail. A pop-up window will show both curves and give details of the measurement result.

Possible reasons for a reference measurement to fail

Optical surfaces (e.g. External reference) dirty or scratched

Environmental conditions out of operational range (e.g. condensating humidity at optical surfaces)

Defective instrument (e.g. electronic damage)

After lamp-module exchange

Remedy

Clean the External reference and the NIRMaster output window

Place the External reference onto the output window and press the ‘Rerun’ button to repeat the

measurement

If measurement still fails, a new reference spectrum can be stored by pressing the ‘Checkmark’
button. A login window will pop-up. An administrator login is required to store the new data set!