Miniav-lt specifications, Caution – Cannon Instrument miniAV-LT Addendum User Manual

Page 2

Advertising
background image

glycol flowing through the hot side heat sinks surrounding the bath
in the Control Unit. Circulation of the mixture helps remove heat
from the Control Unit bath

Filling the Air/Water Heat Exchanger

Obtain a supply of quality automotive antifreeze (ethylene glycol) and
mix it with water in a ratio of 30 percent antifreeze to 70 percent water.
Do not mix antifreeze types in the Air-Water Heat Exchanger.

After you have secured the Exchanger tubing connections (see previous section)
pour this antifreeze/water mixture into the reservoir opening on the top of the
Air/Water Heat Exchanger until it is full (< 2 liters).

CAUTION

Ethylene glycol is a toxic substance. Use proper safety precautions when
handling. Follow appropriate MSDS instructions.

NOTE

Depending on the amount of fluid displacement in the coolant lines, it may be
necessary to add additional antifreeze/water mixture to the Exchanger when the
unit commences operation. You will be able to observe the liquid circulating
inside the reservoir from the opening at the top of the Exchanger during normal
operation. If air bubbles are consistently visible in the coolant lines, add
additional mixture until the Exchanger is full.

CAUTION

The water-antifreeze mix should be replaced annually for reliable performance and to prevent corrosion
of internal components (see
Flushing and draining the Air/Water Heat Exchanger,

miniAV-LT Specifications:

INSTRUMENT SPECIFICATIONS

miniAV-LT Dimensions

BATH UNIT

AIR WATER HEAT EXCHANGER

Dimensions

254 mm wide x 437 mm deep x 598 mm high (10 x
17.25 x 23.5 in)

385.6 mm wide x 511.7 mm deep x 348.8 mm high
(15.2 x 20.15 x 13.7 in)

Weight

27 kg (45 lbs)

14 kg (30 lbs)

Shipping Weight

72 kg (159 lbs) eith all units/accessories

N/A

Operating Conditions

15°C-30°C, 10%-90% RH non-condensing; Installation category IIPollution degree 2

Fuse Rating

115V & 100V UNITS: M250V 8A, 1.25 x 0.25"; 230V Unit: M250V 4A, 1.25 X 0.25"

Compliance

CE Mark: Pending

Computer Requirements

Computer not included, Please contact CANNON for specifications

Catalog Number/
Electrical Requirements

9725-A94

115V AC
50/60 Hz, 850 watt main unit

9725-A96

100V AC

50/60 Hz,850 watt main uni

9725-A95

230V AC
50/60 Hz, 850 watt main unis

Air-Water Heat Exchanger 175 watts




Advertising