Installation:

miniAV-LT setup can be accomplished in just a few minutes by
following the instructions in the installation guide provided with the
instrument.


NOTE: Once the machine and Viscpro II have successfully been

installed, a calibration must be performed at the wash
temperature (30°C) and each temperature at which the user
wishes to run samples.

Preparing the miniAV-LT for testing:

Calibrating the miniAV-LT:

Temperature calibration is essential for the proper operation of the
miniAV-LT after installation. The calibration process consists of
sending a temperature and its offset to the instrument, then waiting
for the temperature to equilibrate. Equilibration is complete when
the instrument stays within 0.02°C for 150 seconds. To calibrate
temperature on the miniAV-LT, follow the instructions below:

NOTE:

Depending on security level settings, it may be necessary for
you to log in as a manager to complete the calibration.

1. Click

Configure from the primary display and select Instrument Settings

from the Configure options for the desired instrument

2. Click

Tray Settings :Tube and Bath option

3.

Enter the temperature you wish to calibrate in the Bath Temp input box
and then press OK to save new target temp

4. Click

Service from the primary display and select your desired

instrument

5. Click

Temperature Calibration option. The Temperature Calibration Data

window will open


NOTE: The value in the Calibration Temperature box should display

the temperature which the user selected in the tray settings

6. Click

Go to Calibration Temperature button to set the target temperature

of the machine to the desired temperature to calibrate (if not the wash
temperature)


NOTE: The miniAV-LT must always be calibrated at 30°C because that

is the wash/idle temperature at which the instrument will
spend most of its time

7.

Place your reference thermometer inside the bath

8.

Wait for the machine to equilibrate at the desired temperature

9.

Enter the current temperature reading to the nearest 0.01°C from the