background image

Two different types of OPTO-RS plug connections are available: Simplex and Duplex

Simplex cable

First generation of OPTO-RS cable connection, designed for instruments which were not able to receive RS232
commands. Data requests are made by LED status change (e.g. by turning off the DTR signal line for a
minimum of 110ms). The OPTO-RS simplex cable can also be used with duplex instruments, however remote
commands will be ignored.
The simplex cable can be directly connected to any standard program as "HyperTerminal" provided with
Windows.
Connections

Line definition

Name

Sub-D 9 pin

Cable color

Line status

Positive power supply :

RTS

7

white

ON (HIGH)

Negative power supply

TXD

3

brown

OFF (LOW)

Data (instrument to periphery)

RXD

2

yellow

INPUT

Data request: Standard status Data
request

DTR

4

green

ON (HIGH) OFF (LOW) during
min. 110m sec.


Duplex cable

The duplex cable allows a 2-way communication between an instrument and a PC in half-duplex mode (e.g. 2-
way communication but not simultanuously).
Important : Only Duplex instruments have the ability to receive RS232 commands. If you use a Duplex cable
with a simplex instrument, all commands other than "?" will be seen as a data request.
The pin assignment of a duplex cable is different to the one of a simplex cable.

Connections

Line definition

Name

Sub-D 9 pin

Cable color

Line status

Positive power supply :

DTR

4

white

ON (HIGH)

Negative power supply

RTS

7

brown

OFF (LOW)

Data (instrument to periphery)

RXD

2

yellow

INPUT

Data request:

TXD

3

green

"?" + <CR>


Note :

In case of data sending from the instrument, the hold mode will be active. To disable the Hold
mode simply do a new data request.

Remote

commands


Format

[¦ C1-Cn ¦ { S1-Sn } ¦ CR ]

C1-Cn:

command of 2 to 3 characters

S1-Sn:

0/1 : command disabled/activated

? : status request

+XXX.YYY: entering numerical values