Parameter:

-In menu [connect to], select[connection using] Directed to {n}
– In menu [parameter], select[terminal keyboard] Suppr. [emulation] ANSI
-In menu [port parameters], select :

[bits/sec]

4800

[data bits]

7

[parity]

Even

[stop bits]

2

[Flux control]

None

Specifications

Connection ......................................................... RS232 compatible, Dsub 9p female or open
Power supply...................................................... from periphery, with TXD, DTR and RTS lines
Data transmission parameters ............................. 4800 bds, even parity, 7 data bits, 2 stop bits
Max. cable length................................................ 15 m according to IEC standards
Number of transmissions /sec.............................. 4-8/sec (depends on the instrument connected)
Data transmission format..................................... [ Sign ¦ E1-En ¦ "." ¦ F1-Fn ¦ CR ]

[ "ERR" ¦ Number ¦ CR ]