GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY

The Compex stimulator is intended for use in the electromagnetic environment specified below. The customer or the user of

the Compex stimulator should assure that it is used in such an environment.

IMMUNITY TEST

IEC 60601

TEST LEVEL

COMPLIANCE LEVEL

ELECTROMAGNETIC ENVIRONMENT -

GUIDANCE

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

1.4 GHz to 2,7 GHz

10 V/m

26MHz to 1GHz

3Vrms

3V/m

10V/m

Portable and mobile RF communications

equipment should be used no closer to any

part of the Compex stimulator, including

cables, than the recommended separation

distance calculated from the equation

applicable to the frequency of the transmitter.

Recommended separation distance

d = 1.2 √P

d = 1.2 √P 80 MHz to 800 MHz

d = 2.3 √P 800 MHz to 2,5 GHz

where P is the maximum output power rating

of the transmitter in watts (W) according to

the transmitter manufacturer and d is the

recommended separation distance

in metres (m).

Field strengths from fixed RF transmitters,

as determined by an electromagnetic site

surveya, should be less than the compliance

level in each frequency rangeb.

Interference may occur in the vicinity of

equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection

from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,

amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess

the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the

measured field strength in the location in which the Compex stimulator is used exceeds the applicable RF compliance level

above, the Compex stimulator should be observed to verify normal operation. If abnormal performance is observed, additional

measures may be necessary, such as reorienting or relocating the Compex stimulator.