Hardy HI 6500 Series User Manual

Page 7

Advertising
background image

 
There are two simple rules for the IP Address: 
 It must have the same network 

identifier as the computer 

 It must have a different node 

identifier than the computer.  
 

5) On a PC running windows, open Internet Properties (TCP/IP).   
6) Click in the ‘Use the Following IP Address’ checkbox; then enter the 

following into the TCP/IP Properties dialog.         

IP Address = 192.168.0.101 

 

Subnet Mask = 255.255.255.000 

7) Select OK on the TCP/IP Properties dialog box. The computer is now 

fully configured. To return the computer to the original network 
settings, return to the ‘Internet Properties (TCP/IP) dialog, select 
‘Obtain an IP address automatically,’ and click OK. 
 

8) The HI 6500 series instrument is now configured to communicate 

with the PC. Enter the HI 6500 series instrument IP address into the 
Windows PC Web browser to access the embedded web browser.   
For example: http://192.168.000.100 

See the HI 6500 User’s Guide for setting up EtherNet/IP,  Ethernet‐UDP 
and Modbus‐RTU communication protocols.  Set up of the Analog 
communications (HI6510 only), Printer, and USB are also located in this 
section.  

Diagnostics:

 

Section used for diagnostics of system. Refer to User’s 
Guide if troubleshooting and use of diagnostics menu is 
required. 

Display:

              Section used for adjusting various aspects of the display,  

including contrast, brightness,  and split screen feature. 

Filter: 

Section used to stabilize weigh readings in an unstable 
environment.  

WAVERSAVER:  Press the enter button and use the arrow keys to select 

the appropriate setting.  WAVERSAVER eliminates 
unwanted noise and is typically set at 1.0 Hz. 

NumAverages:  Set the number of readings to average to provide a 

stable weight.

 

 