KEYENCE LK-G5000 Series User Manual

Page 16

Advertising
background image

16

2. For the calculation method for OUT1, set the measurement target to “[1st peak] – [2nd peak]”.

3. Measurement value scaling

4. Click the button for “Send settings to controller”. (pg. 3 of this guide)

Waveform for received light

from the top surface

Waveform for received light

from the bottom surface

Displaying received light waveforms

Received light waveform when measuring transparent glass

Top surface (1st peak)

Bottom surface (2nd peak)

Master workpiece

(For example, a thickness of

0.1 mm)

Sensor head

Light-receiving

lens

Top surface

reflected light

Bottom surface

reflected light

Transparent glass

Light-receiving

element

(CMOS)

(1) Click “OUT1 Scaling, Offset”.
(2) Click “Set from measured data” in scaling settings.
(3) Measure the master workpiece.
(4) Click “Set” for input value 2.
(5) Enter the thickness of the master workpiece for

Display2. For example, if the thickness of the master
workpiece is 1.0 mm, enter “1.000” for Display2.

(6) Click [OK] and then click “Send settings to controller”.

Click the “Measurement value acquisition start” button and Measurement Value Acquisition/Display will appear.