4 Input/Output Terminals

Sample hold (trigger 2)

• The measurement value is determined after the sampling period (T4) has elapsed by using the tim-

ing input ON as a trigger.

• The minimum input time (T1), input response time (T2), sampling period (T4), and program No.

change time (T7) vary depending on the settings.

• The STROBE time (T3) can be changed.

• Auto-zero is canceled when the ZERO input is turned on for two seconds or more.
• If you turn on the RESET input or the ZERO input during the sampling period, the sampling stops

and the comparator standby state is established.

Refer to page 4-17.

Refer to page 3-43.

Comparator output

ON

OFF

ON

OFF

ON

OFF

ON

OFF

ON

OFF

ON

OFF

Binary output

STROBE output

TIMING input

RESET/

ZERO input

P1 to P3 inputs

0.2ms

Min

MinT1

MinT1

MinT1

T1

MaxT2

MaxT2

MaxT7

MinT1

Min T1

period (T4)

Sampling

T3