Communication performance and, Communication mode in the, Measurement state -3 – KEYENCE LK-G Series User Manual

Page 133: Overview of the settings according to, External devices -3

Advertising
background image

5-3

5

LK-G-M-NO5-E

5 RS-232C

Communication Performance and Communication Mode in the Measurement State

Complete the series of input or output of commands before moving to the tolerance or
function settings from the measurement state.

Overview of the Settings According to External Devices

Perform "Auto transmission settings" according to the external device to be connected as
shown in the following table.

* "Outputting Measurement Values and Changing Settings through Commands" is also valid when

Auto transmission is ON.

*

Refer to "Setting the Communication Specifications of the RS-232C" (page 3-49) for

setting the communication specifications.

Normal
mode

During
measurement:

Receives the measurement control command.
Commands such as for writing or reading the setting values are
not accepted.

During tolerance/
function setting

Communication is not accepted.
The measurement stops.

Communicati
on mode

• [Upon receiving the command [Q], [0] or [CR], the mode

changes to the communication mode, and the setting values
are written or read.

• The measurement stops during the communication mode.

Refer to "Setting the Communication Specifications of the RS-232C" (page 3-49).
For details on usage, refer to the page given in the table.

External device

Auto transmission

Reference page

Non-procedural communication
with a PC and a PLC link unit (the
communication command is
required.)

OFF

"Outputting Measurement Values and
Changing Settings through Commands"
(page 5-4)

Communication with a PC or a
PLC link unit (the measurement
value is outputted by being
synchronized with the external
timing.)

ON

"Outputting Measurement Values in
External Synchronization"
(page 5-22)

Note

Advertising