3

5 Frame ground

FG

4 Signal ground

SG

3 Data set ready

DSR

2 Transmit data

TXD

1 Receive data

RXD

Terminal block J1

4 Relay output common COM

3 Relay output

LO

2 Relay output

OK

1 Relay output

HI

Terminal block J2

D-Sub 9 Pin Connector

D-Sub 9 pin Connector

Connections

SEC-03, Terminal block J1

Pin

Signal name

Pin

Signal name

1 DCD

2 RxD

2

TxD

3 TxD

1

RxD

4 DTR

5 SG

4

SG

6 DSR

3

DSR

7 RTS

8 CTS

Shorted

9 -

Blank heat shrink tubing covered wire

5

FG


AX-KO3341-320










3. SCE-03 RS-232C Serial Interface

• The RS-232C serial interface has four output modes; Stream mode, Command mode, Print switch

mode and Auto-print mode. For details, refer to “

3.3. Data output mode”.

• Set the parameters of the functions “ bp5 ”, “ prt ” and “ aCk ” as necessary.

3.1. Interface specifications

• Data format Baud rate:

2400, 4800, 9600 bps

Data bits:

7 bits + parity 1 bit (even)

Start

bit:

1

bit

Stop bit:

1 bit

Code:

ACII

Terminator:

C

R

L

F

(

C

R

: 0Dh,

L

F

: 0Ah)

LSB

0

1

2

3

4

5

MSB

6






• Relay output maximum rating

Maximum

voltage:

50V

DC

Maximum current:

100mA DC

Maximum

ON

resistance:

8Ω


















3.2. Data format

S T , + 0 0 0 0 0 . 0 0 _ k g C

R

L

F


• Four types of headers are available:

ST : Stable weight data

US : Unstable weight data (including counting data)

QT : Stable counting data

OL : Out of weighing range

• The data is always 9 digits including a sign and a decimal point.

• Four units are available:

_kg : Weight data in “kilograms” (kg)

_lb : Weight data in “decimal pounds” (lb)

_PC : Counting data in “pieces” (pcs)

_oz : Weight data in “decimal ounces” (oz)

• The terminator is always C

R

L

F

.

• Example of output data:

Weight data “kg”

S T , + 0 0 1 2 3 . 4 5 _ k g C

R

L

F

Counting data “pcs”

Q T , + 0 0 0 1 2 3 4 5 _ P C C

R

L

F

Out of weighing range “kg” (+)

O L , + 9 9 9 9 9 . 9 9 _ k g C

R

L

F

Out of weighing range “pcs” (-)

O L , - 9 9 9 9 9 9 9 9 _ P C C

R

L

F


3.3. Data output mode

Stream mode / Command mode Function

setting

prt 0 ”

The scale outputs the current display data in sync with the display update of approximately 10 times per

second. The scale does not output data while it is in the setting mode.

Command mode only

Function setting “ prt 1

The scale replies only to a command

. For details, refer to “

3.4. Command mode”.

1 (-15V~-5V)

0 (5V~15V)

Stop bit

Parity bit

Data bits

Start bit

Header

Data

Unit

Terminator

Separator

Note: “_” indicates a space.

1.Brown

2.Red

3.Orange

4.Yellow

5.Green

6.Blue

7.Purple

8.Gray

9.White

10.Black

RXD

TXD

DSR

SG

FG

SCE-03

Terminal block J1

Optional cable unit

to connect with

AD-8121B printer (3 m)