Safety instructions

Before operation read the safety instructions in the instrument manual (included on

CD as PDF-file). No claims of product liability or warranty can be honoured, if the

device is not operated according to the manual.

If you pass this instrument to somebody else, make sure to include these instructions.

Any questions?

Please contact BYK-Gardner Customer Service - www.byk.com.

DANGER injuries possible

Defects and extraordinary loads.

If safe operation can no longer be presumed, shut down the device and secure it

against unintended operation.

The device must be presumed unsafe to operate:

if visible damage is evident,

if the instrument is no longer working,

if it has been stored for long periods under adverse conditions,

after harsh treatment during shipping.

Safety advices for batteries:

a) Do not crush or dismantle, do not heat or incinerate, do not immerse in any

liquid. This may cause explosion or release harmful substances.

b) Ensure batteries have no short circuit on the contacts (for example due to

damaged insulation). Metallic objects must not come in contact with the bare

contacts.

Do not perform any repairs on the unit.

The unit must be opened by BYK-Gardner authorized repair locations only.

How to disconnect the instrument from any power source:

a) by removing the battery compartment and/or

b) by disconnecting the interface cable

How to disconnect the docking station from any power source:

a) by disconnecting the USB-interface cable and/or

b) by disconnecting the power supply unit plug from the mains socket.

Use only the power supply included with delivery.

WARNING severe material damage

Only devices meeting the requirements for low voltage safety may be connected

to the USB interface.

Additional information on battery disposal

Rechargeable batteries are special waste. Follow the disposal instructions of the

battery or rechargeable battery manufacturer.

Technical Data

Temperature range

10 °C to 42 °C (50 °F to 110 °F) for operation

0 °C to +60 °C (32 °F to 140 °F ) for storage

Rel. humidity

Up to 85% non-condensing/35°C (95 °F)

Interface

USB-B

Accu operation

7.2 V

, max. 1 A

Battery operation

4 AA alkaline batteries (4 AA NiMH batteries optional)

External power supply

Docking station

Input

90 - 264 V

, 50 - 60 Hz, max. 0.8 A

Output

12 V

, 2.5 A