Emission test, Compliance level, Electronic environment guidance – American Diagnostic Corporation (ADC) Diagnostix 2100 User Manual

Page 6: Immunity test

Advertising
background image

The monitor cannot be powered
on.

Indication is suddenly off.

“Err 3” or “Err 4”

“Err 6”

“Err 7”

1. No battery or low battery
power.
2. Battery not installed correctly.
3. The monitor may be damaged.

1. The Oximeter is automatically
powered off, when no signal was
detected after 8 seconds.
2. The battery power is too low to
operate.

1. ERR 3 means the red emission
LED is damaged.
2. Err 4 means the infra-red
emission LED is damaged.

Err 6 means crystal damage.

Err 7 means all the emission LED
or reception dioxide is damaged.

1. Replace battery.
2. Remove and reinstall battery.
3. Contact customer service center.

1. Normal.
2. Replace batteries.

1. Check the red emission LED.
2. Check the infra-red emission LED.
3. Return to Service Center

1. Return to Service Center

1. Return to Service Center

Guidance and Manufacturer’s Declaration: Electromagnetic Emissions

Guidance and Manufacturer’s declaration - electromagnetic environment specified below.

The customer or the user of the PULSE OXIMETER should assure that it is used in such an environment.

Emission Test

RF Emissions CISPR11

RF Emissions CISPR11

Harmonic emissions
IEC 61000-3-2

Voltage fluctuations/

Flicker emissions

IEC 61000-3-2

Compliance Level

Group 1

Class B

Not applicable

Not applicable

Electronic Environment Guidance

The PULSE OXIMETER uses very low RF energy
only and is not likely to cause interference to nearby
electronic equipment.

The PULSE OXIMETER is suitable for use in all es-
tablishments,including domestic and those directly
connected to the public low-voltage power supply
which powers buildings used for domestic pur-
poses.

Immunity

Test

Electrostatic Dis-
charge (ESD)
IEC 61000-4-2

Power Frequency
(50/60 Hz) Mag-
netic Field
IEC 61000-4-8

IEC 60601

Test Level

+/- kV contact
+/- 8kV air

3A/m

Compliance

Level

+/- kV Contact
+/- 8kV air

3A/m

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

For all Equipment and Systems

Guidance and Manufacturer’s declaration - electromagnetic environment specified below.

The customer or the user of the PULSE OXIMETER should assure that it is used in such an environment.

Electronic Environment

Guidance

Floors should be wood, concrete, or ceramic tile.
If floor is covered with synthetic material, the rela-
tive humidity should be at least 30%.

Power frequency magnetic fields should be at lev-
els characteristic of a typical location in a typical
commercial or hospital environment.