Disposable Specula Cleaning/Disinfection Instructions

Disposable specula should not be cleaned or disinfected in any way. Disposable specula are single

patient use only and must be discarded after use. Disposable specula should not be used if they ap-

pear to be visibly contaminated or have accidentally come into contact with contaminated materials

prior to use.

Troubleshooting

The following guidelines should be used to determine if your device has reached its end of life or re-

quires servicing. If the corrective actions described in this section do not resolve your issues, please

see the warranty section of this manual to have your device serviced.

Otoscope and Accessories Troubleshooting Guide

Trouble Area Possible Cause Corrective Action

Lamp burned Replace the lamp with

out ADC part #5111N-4

No light output Wrong lamp Replace the lamp with

installed ADC part #5111N-4

Batteries are Replace with fresh
depleted “AA” batteries

Instrument head Thread/connection Service

will is required

not connect damaged

Wrong specula Use only Welch Allyn brand

Specula will not used or compatible

attach properly

Specula mount is damaged Service required

Loose or misaligned Lens mount Service

magnifying lens is damaged required

Lens is dirty Clean with lint free cloth

Poor or obstructed view

through lens Lens is damaged Service required

Lamp near end Replace the lamp with
Dim light of useful life ADC part #5111N-4

output

Batteries are Replace with fresh
depleted “AA” batteries

Visible corrosion on Damage from excessive Service

instrument or accessories moisture required