Ophthalmoscope Troubleshooting Guide

Trouble Area Possible Cause Corrective Action

Aperture dial is Rotate Aperture dial

between positions

Lamp Burned Replace the lamp with

No light output out ADC part #5112N-4

Wrong Lamp Replace the lamp with

Installed ADC part #5112N-4

Batteries are depleted Replace with fresh “AA” batteries

Instrument head Lamp not fully inserted Re-insert lamp

will

not connect Thread/connection is damaged Service required

Spot is not The aperture dial is not centered Move aperture dial to

centered the full detent position

Lamp near end of useful life Replace the lamp with

Dim light ADC part #5112N-4

output

Batteries are depleted Replace with fresh “AA” batteries

Aperture or lens wheel will not Dirt in or damage Service

turn or does so with difficulty to mechanism required

Visible corrosion on instrument Damage from excessive moisture Service required

Environment

Transportation / Storage: -4°F to 120.2°F (-20°C to +49°C)

RH 95 %

500 hpa – 1060 hpa, Altitude

Environmental Operating Range: 50°F to 120.2°F (10°C to 49°C)

RH 95 % (Max)

500 hpa – 1060 hpa, Altitude

Maximum Operating Temperature:

95°F (35°C)

Symbols

Attention: Read Operating Manual for Cautions and Instructions for Use.

Attention: This is NOT an AP or APG apparatus.

TYPE B. Indicates this is a product with Type B applied parts.