Eu regulatory information, Fcc regulatory information – American Diagnostic Corporation (ADC) ADview Modular Diagnostic Station User Manual

Page 65

Advertising
background image

55

Recommended separation distances between portable and mobile RF communications equipment

and the ADview device

The ADview device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the ADview device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the ADview device as recommended below, according to the
maximum output power of the communications equipment.

Rated maximum

output power of

transmitter

W

Separation distance according to frequency of transmitter

m

150 kHz to 80 MHz

d = [3.5/V

1

] √P

80 MHz to 800 MHz

d = [3.5/V

1

] √P

800MHz to 2.5GHz

d = [7/E

1

] √P

0.01

0.12

0.12

0.23

0.10

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1—At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2—These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.

For units that are Bluetooth enabled, the following EU and FCC regulatory information
applies:

EU Regulatory Information

The transmitter module manufactured by Mitsumi and incorporated into the AdView BP module marked by
has been

approved in accordance with the R&TTE directive.

FCC Regulatory Information

Advertising