2180A

Table 1-4. General Specifications (cont)

OPERATING TEMPERATURE

0 to 50°C

(Y2009: Oto 40°C)

TEMPERATURE COEFFICIENT

±15 ppm/°C from 25°C

ACCURACY vs WIRE LENGTH

No lead resistance error 4-wire RTD's if R

2

adjustment on the RTD Input Module Is
used. Otherwise, 0.04°C/i2 resistance of any

RTD EXCITATION CURRENT

Nominal 0.5 mA.

one input lead if R

2

is not adjusted.

COMMON MODE VOLTAGE

DISPLAY

F or C, switch-selectable, 7-segment LEDs 1.1 cm

high

350V dc, 250V ac rms max.

NORMAL MODE REJECTION

RESOLUTION

0.01 below 204 C for platinum RTD's, automatically

shifting to 0.1° above 204°C (93°C for nickel RTD's).

If readings are decreasing, shift Is at 77°C for platinum,

66°C for nickel.

>90 dB at DC, 50, 60, and 400 Hz ±0.1%.

RESPONSE TIME

1 second typical.

MEASUREMENT METHOD

ZERO DRIFT

Dual slope integration, under microcomputer control.

100 ms integration period, three readings per second.
A/D Resolution is 100,000 counts at full-scale.

Automatic zero correction.

RTD TYPES

LINEARIZATION TECHNIQUES

Computer algorithm, 4th order curve fit.

100S2 385 Pt (DIN), 390 Pt, 3916 Pt, 392 Pt,

100S2NÌ (DIN), lOnCU.

COMMON MODE REJECTION

>160 dB at DC, 50, 60, and 400 Hz ±0.1% with 100i2
unbalance at inputs.

VOLTAGE RANGE (CALIBRATION ONLY)

No annunciator or decimal point.

99999

m

V full scale (switch S2 in AUTO), resolution 1

¡A/

999990

m

V full scale (switch S2 in .1), resolution 10

m

V

SAFETY

Protection Class 1
Relates solely to insulation or grounding properties

defined in lEC 348.

1-5/1-6